CVE-2019-25722: Dräger SC Monitoring devices contain hardcoded credentials and denial-of-service risk

CVE-2019-25722 · published 5 months ago
Summary

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) store passwords in plain text, making them easily accessible to unauthorized users. This, combined with a potential denial-of-service (device crash) vulnerability, poses a risk to the integrity of patient monitoring and network connectivity. To mitigate this risk, ensure that the devices are updated to the latest version or consider replacing them with secure alternatives.

Original advisory text
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and...
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
Fix within
Internet-facing 60 days
Internal At next upgrade
  • Not known to be exploited
  • Needs hands-on effort to exploit
  • Gives an attacker partial control
Severity
7.2 High
Exploitation
<1% chance of attack within 30 days
Type
CWE-798Use of Hard-coded Credentials
Timeline
Published2 Jun 2026
Updated29 Oct 2026
First seen2 Jun 2026
Sources
Track software like this