Track vulnerabilities like this one. Sign up free to get alerted when software you use is affected.
8.1

CVE-2017-12724: Medfusion 4000 pump FTP server allows default login

CVE-2017-12724 · published 8 years ago
Summary

The Medfusion 4000 wireless syringe infusion pump includes an FTP service that ships with a preset username and password. If the device is set to permit FTP connections, anyone who knows these default credentials could access the pump's file system. Disable FTP access or change the credentials to a strong, unique password to remove the risk.

What to do

The CVE record does not list a fixed version. Check the vendor's site or the advisory links below - a fix may already be released.

Affected software
VendorProductAffected versions
smiths-medical medfusion_4000_wireless_syringe_infusion_pump 1.1
1.5
1.6
cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.1:*:*:*:*:*:*:*
Original advisory text
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded creden...
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.
References
Fix within
Internet-facing 3 days
Internal 60 days
  • Not known to be exploited
  • Can be exploited automatically (estimated)
  • Gives an attacker full control (estimated)
Severity
8.1 High
Exploitation
1% chance of attack within 30 days
Type
CWE-798Use of Hard-coded Credentials
Timeline
Published15 Feb 2018
Updated4 Nov 2026
First seen29 Sep 2026
Sources
Track software like this
Free during beta